Improving Electrocardiogram Diagnostic Reliability in Viet Nam: Electrode Placement, Signal Quality and Interpretation Safeguards

DIAGNOSTIC SAFETY & RISK MANAGEMENTMEASUREMENT, LEARNING & IMPROVEMENT

11/5/202617 min read

Improving Electrocardiogram Diagnostic Reliability in Viet Nam: Electrode Placement, Signal Quality and Interpretation Safeguards

Two electrocardiograms recorded only minutes apart can look surprisingly different. Sometimes the change reflects a genuine alteration in the patient's cardiac condition. In other cases, it may result from an electrode placed in the wrong position, a reversed cable connection, or interference during recording.

Distinguishing between these possibilities matters. A technical problem may produce an ECG pattern resembling cardiac disease and lead to unnecessary investigations. It may also obscure a clinically important abnormality or complicate decisions about urgent care. Yet an unexpected tracing should not be dismissed as a technical error simply because its findings are difficult to explain.

The standard 12-lead electrocardiogram (ECG) remains one of the most widely used diagnostic investigations in healthcare. It is relatively quick, noninvasive, and available in emergency departments, inpatient units, outpatient clinics, and specialist services. Its familiarity, however, can create the impression that obtaining a reliable recording is straightforward.

An ECG is more than a tracing generated by a machine. Its diagnostic value depends on how accurately electrical signals are acquired, whether the recording is technically suitable for interpretation, and how the findings are evaluated alongside the patient's symptoms, clinical history, and other evidence.

Improving ECG reliability therefore requires more than teaching clinicians to recognize abnormal waveforms. Healthcare organizations also need dependable recording practices, suitable equipment, competent personnel, and practical mechanisms for identifying technical and interpretive problems.

Two investigations published in 2025 by the United Kingdom's Health Services Safety Investigations Body (HSSIB) provide useful lessons. The investigations examined the recording and interpretation of 12-lead ECGs in ambulance services, highlighting concerns involving professional education, electrode placement, equipment usability, automated interpretation, and clinical decision support.

Although these findings arose from emergency medical services in England, they raise questions relevant to hospitals and outpatient facilities elsewhere, including Viet Nam. How consistently are ECGs recorded? Can clinicians recognize when technical conditions may have altered the tracing? And what safeguards prevent an automated statement from being accepted without appropriate professional assessment?

Diagnostic Reliability Begins With Accurate ECG Acquisition

A standard resting 12-lead ECG uses ten electrodes to generate twelve electrical views of the heart. Six electrodes are positioned on the chest, while four are placed on the limbs. Their locations matter because the recorded waveforms reflect electrical activity measured from defined anatomical perspectives.

When an electrode is placed incorrectly, the resulting ECG may differ from the tracing that would have been obtained under standard recording conditions. Changes in chest electrode positioning can alter waveform morphology, R-wave progression, and findings used to assess possible cardiac abnormalities.

These effects may be difficult to recognize from the finished recording alone. An ECG affected by electrode misplacement may still display all twelve leads, produce apparently plausible waveforms, and generate a computerized interpretation. The recording may look complete even though its diagnostic reliability has been compromised.

The American Heart Association, American College of Cardiology Foundation, and Heart Rhythm Society addressed these concerns in their 2007 scientific statement on ECG standardization and technology. Although published nearly two decades ago, the statement remains an important foundational reference for electrode positioning, recording characteristics, signal processing, and computerized interpretation.

More recent publications reinforce the continuing relevance of accurate acquisition. In 2025, Sampson described the practical requirements for obtaining a reliable 12-lead ECG, including appropriate skin preparation, electrode positioning, and prevention of cable connection errors.

For healthcare organizations, the central question is whether an ECG was obtained under conditions sufficiently reliable for its intended clinical use. Successful generation of a tracing is not, by itself, evidence that the acquisition process was correct.

Electrode Placement Can Change the Meaning of a Recording

Accurate positioning of chest electrodes is fundamental to a standard diagnostic 12-lead ECG.

V1 is placed in the fourth intercostal space at the right sternal border, and V2 in the fourth intercostal space at the left sternal border. V4 is positioned in the fifth intercostal space along the left midclavicular line, with V3 between V2 and V4. V5 is placed at the left anterior axillary line and V6 at the left midaxillary line, both at the same horizontal level as V4.

These locations are established anatomical reference points rather than approximate positions selected by visual estimation.

One recognized source of error is positioning V1 and V2 too high on the chest. This can alter the recorded waveforms and produce patterns that resemble cardiac abnormalities, including changes in R-wave progression, T-wave morphology, and other features relevant to interpretation.

In a 2018 publication in the American Journal of Emergency Medicine, Walsh described the potential clinical consequences of incorrect V1 and V2 placement, including patterns that may be mistaken for ischemic changes, conduction abnormalities, or other cardiac conditions.

Incorrect connections between limb electrodes and their corresponding cables can produce another set of misleading findings. Depending on the connections involved, the tracing may show unexpected changes in waveform polarity, electrical axis, or relationships between leads.

These patterns may suggest a technical problem, but they are not conclusive evidence of one. Genuine cardiac abnormalities can also produce unusual ECG appearances.

When electrode misplacement or cable reversal is suspected, checking electrode positions and connections may be appropriate. A correctly obtained repeat recording can help determine whether the unexpected pattern persists.

The decision to repeat an ECG must nevertheless reflect the clinical circumstances. In a patient requiring urgent assessment, technical verification should not cause an avoidable delay in necessary care.

The distinction between diagnostic ECG recordings and cardiac monitoring configurations is also important.

In certain monitoring situations, electrodes normally placed on the limbs may be repositioned on the torso to accommodate movement, patient condition, or practical requirements. Such configurations may be appropriate for their intended monitoring purpose, but they are not necessarily interchangeable with standard diagnostic 12-lead recordings.

The HSSIB investigation published in October 2025 specifically identified the potential effect of torso-based limb electrode placement on ECG waveforms and automated interpretation when recordings were used to support diagnosis.

Alternative electrode positions may be necessary for patients with amputations, movement disorders, injuries, or other physical limitations. Where modifications are required, their implications should be understood and documented appropriately, particularly when comparing recordings obtained at different times.

Accurate electrode placement must also be compatible with respectful patient care. Identifying anatomical landmarks may involve examination of areas covered by clothing or breast tissue. Appropriate explanation, privacy, respect, and adherence to applicable consent requirements are essential.

For healthcare organizations, electrode placement competence should be supported by practical training and observation of actual recording practices, rather than relying exclusively on written instructions or initial orientation.

Signal Quality: Recognizing When a Tracing Cannot Be Trusted

Correct electrode positioning is necessary, but it does not guarantee a technically reliable ECG.

Signal quality may be affected by patient movement, muscle activity, poor electrode contact, inadequate skin preparation, damaged cables, and electrical interference. These problems can produce baseline fluctuations, excessive noise, distorted waveforms, or patterns resembling genuine rhythm abnormalities.

Some artifacts are readily recognized. Others are more difficult to distinguish from clinically important findings, particularly when the tracing is reviewed without information about the circumstances of recording.

Patients with tremors, difficulty remaining still, acute distress, or limited ability to cooperate may present additional challenges. Recording conditions may also be less controlled in emergency settings than during routine outpatient examinations.

Appropriate skin preparation, secure electrode contact, suitable patient positioning, and attention to avoidable sources of interference can improve signal acquisition. Equipment should be maintained according to manufacturer instructions, and recurring problems with cables or connectors should prompt technical assessment.

Recording settings deserve equal attention.

Modern electrocardiographs use digital processing and filtering to reduce unwanted signals. These functions can improve readability, but inappropriate settings may alter clinically important waveform characteristics.

For example, excessive filtering can affect features relevant to ST-segment assessment or reduce the visibility of certain high-frequency components of the ECG.

Settings suitable for continuous cardiac monitoring should not automatically be assumed appropriate for every diagnostic ECG application.

Healthcare organizations should ensure that equipment is configured for its intended use and that personnel understand how recording settings may affect interpretation.

The displayed or stored ECG should preserve relevant technical information, including lead identification, recording time, calibration, and settings needed to interpret or compare tracings appropriately. Paper speed, signal gain, and applicable filter settings may be particularly relevant when measurements or waveform characteristics are compared.

When signal quality is inadequate, repeating the recording under improved conditions may be necessary. However, attempting to obtain a technically perfect tracing should not delay urgent clinical decisions when the patient's condition requires immediate assessment.

The practical objective is to recognize when technical limitations materially affect the reliability of the information and respond in a manner proportionate to the clinical risk.

Automated ECG Interpretation Is Useful but Not Definitive

Many modern electrocardiographs generate automated measurements and interpretive statements. These functions may identify heart rate, rhythm characteristics, conduction intervals, electrical axis, and patterns associated with particular cardiac abnormalities.

Automated analysis can support clinical workflow, especially when clinicians need rapid access to preliminary information. Nevertheless, its limitations should be understood.

The accuracy of computerized interpretation depends on signal quality, electrode placement, the algorithm used, patient information, and the types of abnormalities being assessed.

A statement indicating a normal ECG does not establish that the patient has no cardiac disease. Similarly, a computerized warning suggesting myocardial infarction does not, by itself, confirm the diagnosis.

The second HSSIB investigation, published on October 30, 2025, examined the role of automated interpretation in recognizing suspected ST-elevation myocardial infarction (STEMI) in ambulance services.

The investigation identified differences in algorithm performance, the evidence supporting reported accuracy, and how interpretive statements were presented to users.

Particular concerns arose when ECG changes were borderline or did not clearly meet conventional STEMI criteria. Such cases may require careful professional assessment rather than reliance on a single software-generated conclusion.

The HSSIB findings should not be generalized into one accuracy estimate for every ECG algorithm. Different devices and software products may perform differently, and findings from ambulance services in England cannot be assumed to represent performance in all healthcare settings.

Some computerized interpretation systems incorporate patient age and sex into their analysis. Missing or inaccurate information may affect the generated interpretation, depending on the specific software.

Correct patient identification and data entry are therefore part of reliable ECG acquisition, not merely administrative tasks.

Healthcare organizations should also distinguish conventional computerized interpretation from newer artificial intelligence-based diagnostic applications. Some traditional systems rely primarily on predefined algorithms and rules, while others incorporate machine learning. Their design, validation, intended uses, and limitations are not necessarily identical.

Regardless of the underlying technology, an automated statement should support rather than replace appropriate professional assessment.

The interpreting clinician needs to examine the tracing, consider its technical quality, evaluate the findings in the clinical context, and recognize when further information or specialist advice may be necessary.

Automated interpretation also does not remove the healthcare organization's responsibilities for appropriate clinical oversight, professional competence, and safe equipment use. Manufacturers and software providers may have separate obligations under applicable medical device and product regulations.

A Technically Reliable ECG May Still Be Clinically Inconclusive

One of the most important distinctions in ECG interpretation is the difference between technical reliability and diagnostic sufficiency.

An ECG may be recorded correctly, with adequate signal quality and no apparent electrode placement errors, yet still be insufficient to answer the clinical question.

Consider a patient with ongoing symptoms suggestive of acute myocardial ischemia. The initial ECG may show no diagnostic changes despite the possibility of an evolving acute coronary syndrome.

In this situation, the recording may accurately represent the heart's electrical activity at the time it was obtained. The limitation is not necessarily a technical error. It may reflect the evolving nature of the condition or the diagnostic limitations of a single ECG.

The 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of acute coronary syndromes emphasizes timely ECG acquisition and interpretation when acute coronary syndrome is suspected.

For patients with suspected acute coronary syndrome in the prehospital setting, the guideline recommends acquiring and interpreting a 12-lead ECG within 10 minutes of first medical contact. It also recommends ECG acquisition and interpretation within 10 minutes during the initial hospital assessment of patients with suspected acute coronary syndrome.

These recommendations apply to defined clinical circumstances. They should not be interpreted as universal time targets for every ECG examination or every healthcare service.

The guideline also recommends serial 12-lead ECGs when an initial tracing is nondiagnostic and clinical suspicion remains high, symptoms persist, or the patient's condition deteriorates.

The comprehensive Australian clinical guideline for acute coronary syndromes was published in Heart, Lung and Circulation in 2025, followed by a guideline summary in the Medical Journal of Australia in February 2026.

The Australian guidance emphasizes the importance of recognizing ECG patterns that may indicate acute coronary occlusion, including presentations that do not meet conventional ST-elevation criteria. It also supports additional ECG assessment when symptoms persist, change, or recur, or when diagnostic uncertainty remains.

These recommendations reinforce the need to interpret ECG findings alongside symptoms, clinical assessment, cardiac biomarkers, and other relevant evidence.

An initial ECG without diagnostic changes should not automatically be used to exclude acute coronary syndrome.

Depending on the presentation, further assessment may involve repeat ECGs, comparison with previous recordings, cardiac troponin testing, additional investigations, or specialist evaluation.

Such decisions must follow the relevant clinical pathway and be made by appropriately qualified professionals.

The distinction also matters when comparing ECGs obtained at different times. Apparent changes may reflect genuine disease progression, differences in electrode placement, altered recording settings, or a combination of factors.

When findings are inconsistent, clinicians should consider these possibilities rather than immediately assuming that either the previous or current recording is incorrect.

A reliable diagnostic process must recognize both kinds of uncertainty: a tracing that may be technically compromised and a technically satisfactory tracing that remains clinically inconclusive.

ECG Reliability Requires Coordinated Professional Responsibilities

A dependable ECG process involves several professional roles.

Personnel obtaining the recording need the practical skills to identify anatomical landmarks, prepare the patient, operate equipment correctly, recognize artifacts, and identify situations requiring additional attention.

Clinicians interpreting ECGs need sufficient competence to recognize important abnormalities, assess recording limitations, and integrate findings with other clinical information.

Biomedical engineering and technical personnel contribute through equipment maintenance, troubleshooting, and management of device configuration. Quality and clinical governance teams can support competency assessment, incident review, and monitoring of recurring problems.

These responsibilities overlap, but they are not interchangeable.

For example, a person recording an ECG may recognize an unusual tracing without being authorized or qualified to determine its diagnostic significance. A dependable process should allow that concern to be communicated promptly to an appropriate clinician.

Similarly, when a clinician suspects electrode misplacement after reviewing a tracing, there should be a practical mechanism for checking the recording conditions and obtaining a corrected ECG where clinically appropriate.

These arrangements are particularly important when ECG recording occurs across multiple departments or when personnel have different levels of experience.

Training should extend beyond initial orientation. Practical observation, equipment-specific instruction, periodic competency reassessment, and feedback from reviewed cases can help maintain reliable performance.

Device procurement and replacement also deserve consideration.

The selection of an electrocardiograph should take account of signal quality, usability, display clarity, technical support, compatibility with clinical information systems, and the limitations of automated interpretation.

Acquiring equipment with more functions does not necessarily improve diagnostic reliability if personnel cannot use those functions correctly or the surrounding workflow is poorly organized.

Changes to equipment settings, software versions, or interpretive algorithms should be appropriately managed because they may affect recording characteristics, measurements, displayed interpretations, and clinical workflow.

Professional competence and equipment performance need to be supported by organizational arrangements that allow technical concerns to be recognized and addressed before they compromise patient care.

Measuring ECG Quality Without Drawing Misleading Conclusions

ECG quality improvement requires more than confirming that equipment is available or that recordings have been completed.

Healthcare organizations need measures that help identify technical weaknesses, assess interpretation processes, and determine whether corrective actions are effective.

One useful indicator is the proportion of ECG recording attempts requiring repetition because of technical inadequacy.

However, repeats performed for technical reasons should be distinguished from serial ECGs obtained to assess evolving clinical findings.

A repeat ECG requested because symptoms persist or the patient's condition changes is not evidence that the original recording was performed incorrectly.

Another potential measure is the frequency of confirmed electrode placement errors.

This requires a credible assessment process. Direct observation, documented verification, or other appropriate evidence may establish that an electrode was misplaced. An unusual waveform alone should not automatically be counted as a confirmed technical error.

Signal quality can be evaluated through structured review of selected recordings, with attention to baseline instability, muscle artifact, poor electrode contact, lead problems, and other limitations that affect interpretability.

Hospitals may also examine the completeness of patient identification and recording information, the frequency of equipment-related problems, and the effectiveness of competency assessment arrangements.

Interpretation quality is more difficult to measure.

Disagreement between a computerized statement and a clinician does not necessarily establish that either interpretation is incorrect. Likewise, two qualified clinicians may disagree about an ECG without the difference automatically constituting a diagnostic error.

Meaningful evaluation requires appropriate expertise, predefined review criteria, and a method for considering clinically significant discrepancies.

Some ECGs may be selected for review because they contain unusual findings, were associated with an important clinical decision, or raised questions about technical quality.

Such targeted reviews are useful for learning, but they should not be treated as unbiased estimates of the overall error rate across all ECG recordings.

A broader sample of routine examinations can provide complementary information.

Performance indicators should also be interpreted in context.

A low technical repeat rate may reflect good acquisition practices, but it may also indicate that poor-quality recordings are not being recognized. Conversely, a temporary increase in repeat recordings after introducing quality review may reflect improved identification of technical problems.

No single measure can establish diagnostic safety.

The value of measurement lies in identifying recurring weaknesses, assessing whether corrective actions work, and determining whether reliable ECG acquisition and interpretation are becoming more consistent.

Applying International Guidance Within Viet Nam's Healthcare System

ECG recording is performed across a wide range of healthcare settings in Viet Nam, including tertiary hospitals, provincial hospitals, outpatient facilities, emergency services, and specialist clinics.

The availability of equipment, staffing arrangements, access to specialist interpretation, and maturity of clinical information systems may differ between facilities.

These differences make a proportionate approach to ECG quality assurance important.

A healthcare organization does not necessarily need an advanced electronic monitoring platform to begin improving reliability. Clear recording procedures, practical competency assessment, functioning equipment, and periodic review of selected tracings can provide a useful foundation.

Facilities with more developed information systems may incorporate electronic ECG storage, structured review records, audit trails, and decision-support functions.

These capabilities can improve information availability and traceability, but they do not remove the need for competent recording and interpretation.

International technical and clinical references can inform local procedures.

The AHA/ACCF/HRS ECG standardization statement provides foundational principles for electrode placement, signal acquisition, and interpretation. IEC 60601-2-25:2011 addresses particular requirements for the basic safety and essential performance of electrocardiographs intended for diagnostic use.

These documents serve different purposes.

A technical standard for electrocardiographs does not, by itself, constitute a complete clinical quality assurance program. Likewise, international professional recommendations should not automatically be interpreted as legally mandatory for every healthcare facility in Viet Nam.

The relevant Vietnamese framework includes the Law on Medical Examination and Treatment No. 15/2023/QH15, effective January 1, 2024, as amended and reflected in Consolidated Document No. 26/VBHN-VPQH dated February 26, 2026.

Decree No. 96/2023/ND-CP provides implementing provisions concerning medical examination and treatment activities.

Circular No. 23/2024/TT-BYT, which establishes the list of medical examination and treatment techniques, as amended by Circular No. 25/2026/TT-BYT effective August 15, 2026, is also relevant when organizations determine their authorized technical activities.

Healthcare facilities and professionals must operate within applicable licensing conditions, authorized professional scope, and other relevant legal requirements.

International guidance can inform local quality procedures but does not independently authorize a professional to perform or interpret ECGs outside the scope permitted by Vietnamese law.

Digital ECG management introduces additional considerations concerning patient information.

ECG records may contain identifiable health information, examination dates, diagnostic findings, and other clinical details. When recordings are stored, transmitted, or reviewed electronically, appropriate controls should protect their accuracy, integrity, confidentiality, and authorized use.

Viet Nam's Law on Personal Data Protection No. 91/2025/QH15 and Decree No. 356/2025/ND-CP took effect on January 1, 2026.

Healthcare organizations should assess the lawful basis for processing patient information, applicable consent requirements and exceptions, access controls, information security, and conditions governing data sharing or transfer.

Internal quality assessment, educational use, research, and external consultation may involve different purposes and should be managed according to the requirements applicable to each activity.

A dependable ECG service must preserve clinically useful information while respecting patients' rights and confidentiality.

Turning ECG Quality Review Into Practical Improvement

Quality assurance becomes valuable when it improves what happens during everyday clinical care.

Consider a hospital that repeatedly identifies V1 and V2 electrodes being placed above their intended anatomical positions.

Correcting an affected recording may resolve the immediate concern. Preventing recurrence, however, requires understanding why the placement errors occurred.

Personnel may rely on visual estimation rather than identifying anatomical landmarks. Training may vary between departments. Newly recruited staff may receive limited supervised practice, or recording conditions may make accurate positioning difficult.

Simply instructing personnel to be more careful may have limited effect if these contributing factors remain unchanged.

A focused improvement effort could involve practical retraining, supervised observation, clearer recording instructions, and subsequent reassessment of electrode placement accuracy.

The organization can then examine whether the frequency of confirmed errors has decreased and whether recording quality has improved.

A similar approach applies to recurring signal artifacts, cable connection problems, inaccurate patient information, equipment configuration concerns, and interpretive discrepancies.

Some issues may be addressed through training or workflow adjustments. Others may require technical maintenance, vendor support, or improved access to specialist interpretation.

Feedback should be timely, constructive, and connected to actual clinical practice.

Personnel obtaining ECGs benefit from understanding how technical decisions influence diagnostic interpretation. Clinicians interpreting ECGs benefit from reliable information about the circumstances in which recordings were acquired.

This creates an opportunity for shared learning rather than assigning all responsibility to one professional group.

For healthcare organizations in Viet Nam, the priority should not simply be to adopt the most sophisticated ECG technology available. It should be to establish processes capable of producing dependable recordings, recognizing their limitations, and responding appropriately when technical or interpretive uncertainty arises.

From a Reliable Tracing to Safer Diagnostic Decisions

An ECG may take only a few minutes to record, but the information it provides can substantially influence clinical decisions.

Reliable electrocardiography begins with accurate electrode placement, appropriate recording settings, and adequate signal quality. Its diagnostic usefulness also depends on professional interpretation and recognition that a technically satisfactory tracing may not fully resolve the clinical question.

Automated interpretation can support these processes, but it should not substitute for clinical assessment or create false reassurance.

For healthcare organizations in Viet Nam, strengthening ECG quality does not necessarily require the most advanced equipment. It requires consistent technical practices, competent personnel, appropriate clinical oversight, and meaningful review of recording and interpretation problems.

The most useful quality improvement process helps professionals recognize potential artifacts, identify when additional clinical assessment is justified, and improve ECG recording practices without encouraging unnecessary investigations.

Diagnostic excellence in electrocardiography is not demonstrated simply by obtaining a tracing quickly or generating an automated interpretation. It depends on whether the ECG was recorded reliably, interpreted in the appropriate clinical context, and supported by safeguards that help clinicians recognize uncertainty and make safer decisions.

Selected References and Regulatory Sources

Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram: Part I: The Electrocardiogram and Its Technology. Circulation. 2007;115(10):1306–1324. DOI: 10.1161/CIRCULATIONAHA.106.180200.

Health Services Safety Investigations Body. 12-Lead Electrocardiograms (ECGs) in Ambulance Services: Paramedic Education, Training and Competence. Investigation report. March 27, 2025.

Health Services Safety Investigations Body. 12-Lead Electrocardiograms (ECGs) in Ambulance Services: Diagnosis of Suspected ST Elevation Myocardial Infarction (STEMI). Investigation report. October 30, 2025.

Rao SV, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI Guideline for the Management of Patients With Acute Coronary Syndromes. Circulation. 2025;151(13)–e862. DOI: 10.1161/CIR.0000000000001309.

Brieger D, Cullen L, Briffa T, et al. National Heart Foundation of Australia & Cardiac Society of Australia and New Zealand: Comprehensive Australian Clinical Guideline for Diagnosing and Managing Acute Coronary Syndromes 2025. Heart, Lung and Circulation. 2025;34(4):309–397. DOI: 10.1016/j.hlc.2025.02.102.

Brieger DB, Cullen LA, Briffa TG, et al. National Heart Foundation of Australia and Cardiac Society of Australia and New Zealand: Australian Clinical Guideline for Diagnosing and Managing Acute Coronary Syndromes 2025. Medical Journal of Australia. 2026;224(2). DOI: 10.5694/mja2.70127.

Sampson M. How to Accurately Record a 12-Lead ECG. Nursing Standard. 2025;40(8):49–54. DOI: 10.7748/ns.2025.e12512.

Walsh B. Misplacing V1 and V2 Can Have Clinical Consequences. American Journal of Emergency Medicine. 2018;36(5):865–870. DOI: 10.1016/j.ajem.2018.02.006.

International Electrotechnical Commission. IEC 60601-2-25:2011. Medical Electrical Equipment, Part 2-25: Particular Requirements for the Basic Safety and Essential Performance of Electrocardiographs. Second edition. 2011.

National Assembly of Viet Nam. Law on Medical Examination and Treatment No. 15/2023/QH15. January 9, 2023. Effective January 1, 2024, as amended. See also Consolidated Document No. 26/VBHN-VPQH, dated February 26, 2026.

Government of Viet Nam. Decree No. 96/2023/ND-CP, Detailing Certain Provisions of the Law on Medical Examination and Treatment. December 30, 2023, as amended where applicable.

Ministry of Health of Viet Nam. Circular No. 23/2024/TT-BYT, Promulgating the List of Medical Examination and Treatment Techniques. October 18, 2024, as amended.

Ministry of Health of Viet Nam. Circular No. 25/2026/TT-BYT, Amending and Supplementing Certain Provisions of Ministry of Health Circulars, Including Circular No. 23/2024/TT-BYT. June 30, 2026. Effective August 15, 2026.

National Assembly of Viet Nam. Law on Personal Data Protection No. 91/2025/QH15. Effective January 1, 2026.

Government of Viet Nam. Decree No. 356/2025/ND-CP, Detailing Certain Provisions and Implementation Measures of the Law on Personal Data Protection. Effective January 1, 2026.

Professional Scope and Disclaimer

This article is published by Diagnostic Excellence Vietnam (DxEx.vn), a professional initiative dedicated to advancing diagnostic quality, safety, and reliability in Viet Nam through evidence-informed knowledge sharing, risk management, measurement, and quality improvement.

The content is intended for general professional education and discussion. It does not constitute patient-specific medical advice, a clinical diagnosis, an individualized ECG recording or interpretation protocol, a recommendation for particular diagnostic investigations or treatment, a determination of professional or technical error, an accreditation assessment, or legal advice.

The discussion focuses primarily on standard diagnostic 12-lead electrocardiography. Its recommendations should not be applied without appropriate adaptation to continuous cardiac monitoring, ambulatory ECG monitoring, exercise testing, pediatric electrocardiography, or other specialized recording applications.

References to international technical standards, clinical guidelines, professional recommendations, and research findings do not imply endorsement of this publication by the organizations concerned or that these materials are automatically legally mandatory in Viet Nam.

Decisions concerning ECG acquisition, electrode placement, interpretation, repeat recordings, further investigations, urgent care, and subsequent management must be made by appropriately qualified and authorized healthcare professionals in accordance with applicable Vietnamese law, relevant professional standards, authorized service capabilities, and the patient's clinical circumstances.