Improving Diagnostic Ultrasound Reliability in Viet Nam: Examination Completeness, Image Documentation and Quality Review

DIAGNOSTIC PATHWAYS & RELIABILITYMEASUREMENT, LEARNING & IMPROVEMENT

11/5/202624 min read

Improving Diagnostic Ultrasound Reliability in Viet Nam: Examination Completeness, Image Documentation and Quality Review

A patient undergoes abdominal ultrasound because of persistent pain. The report describes no significant abnormality, and several images are stored in the hospital's imaging system. When the patient returns for further assessment, another clinician reviews the examination but cannot determine whether all the relevant structures were adequately visualized. Some expected images are missing, and the report does not explain whether technical limitations affected the findings.

The original interpretation may have been reasonable. Missing archived images do not prove that the examination was performed incorrectly, and a later diagnosis does not automatically establish that an abnormality was overlooked. Nevertheless, incomplete documentation makes it difficult to understand what was actually assessed and how much confidence should be placed in the reported findings.

This is a particular challenge in diagnostic ultrasound. Unlike imaging examinations that acquire a relatively standardized dataset for subsequent interpretation, ultrasound relies heavily on decisions made during real-time scanning. The person performing the examination selects scanning planes, adjusts equipment settings, evaluates anatomical structures, and determines whether additional views are needed. The images stored afterward may represent only a portion of what was observed.

Diagnostic reliability therefore depends on more than technically acceptable images or a well-written report. It requires an examination appropriate to the clinical question, adequate visualization of relevant structures, meaningful image documentation, and an effective process for reviewing examination quality.

Recent international guidance provides an updated foundation for these activities. The American Institute of Ultrasound in Medicine (AIUM) published revised documentation guidance in January 2026, updated its ultrasound practice accreditation standards in April, and released a revised practice parameter for abdominal and retroperitoneal ultrasound in September.

For healthcare organizations in Viet Nam, these developments offer an opportunity to examine whether ultrasound services consistently produce information that is clinically relevant, appropriately documented, and sufficiently reliable to support subsequent decisions.

The objective is not to require identical images for every patient or assume that every deviation from a standard protocol represents poor care. It is to make the intended scope of the examination clear, recognize important limitations, and preserve enough information to support interpretation and quality review.

What Makes a Diagnostic Ultrasound Examination Reliable?

Diagnostic ultrasound uses high-frequency sound waves to provide information about anatomical structures and selected physiological or pathological changes. Its advantages include real-time visualization, assessment of certain dynamic findings, and the absence of ionizing radiation.

These advantages do not eliminate technical and interpretive limitations. Image quality can be affected by body habitus, bowel gas, anatomical variations, patient positioning, restricted mobility, and the availability of suitable acoustic windows. The choice of transducer, scanning plane, imaging depth, gain, focal settings, and other technical parameters can also influence how well relevant structures are visualized.

Several aspects of examination quality need to be considered together.

Technical adequacy concerns whether the examination produced images of sufficient quality for the intended assessment. Examination completeness concerns whether the relevant anatomical structures and required components were appropriately evaluated. Diagnostic adequacy concerns whether the information obtained was sufficient to support the interpretation, including recognition of material limitations.

Documentation quality concerns whether the stored images and final report adequately represent the examination and allow relevant findings to be understood and reviewed.

These aspects are closely related, but they are not interchangeable.

An examination may be technically well performed within a focused scope without being sufficient to answer a broader clinical question. Another may cover the intended anatomical structures but remain diagnostically limited because important areas could not be visualized adequately.

Similarly, a collection of technically excellent images does not necessarily demonstrate that every relevant component of the examination was performed. A written report may accurately describe what was seen while providing insufficient information about what could not be assessed.

Recognizing these differences helps healthcare organizations identify the actual source of a quality concern. Inadequate visualization may require technical adjustments, different equipment, or modified patient preparation. Incomplete anatomical coverage may indicate a need for clearer protocols or additional training. Weak documentation may require improvements in image capture, labeling, reporting, or archival procedures.

Effective quality review should identify where the process needs improvement rather than treating every ultrasound concern as the same type of problem.

Examination Completeness Depends on the Clinical Question

Not every ultrasound examination needs to be comprehensive. The appropriate scope depends on the patient's presentation, the clinical question, and the type of examination requested.

A comprehensive abdominal ultrasound may involve evaluation of several abdominal and retroperitoneal structures. A focused examination may be directed toward a particular organ, anatomical region, or previously identified abnormality. Both approaches can be appropriate when their purpose and limitations are understood.

The AIUM Practice Parameter for the Performance of an Ultrasound Examination of the Abdomen and/or Retroperitoneum, 2026 Revision, published on September 11, 2026, recognizes this distinction.

The parameter explains that abdominal ultrasound may evaluate the entire abdomen and retroperitoneum, selected organs, or a more focused anatomical region, depending on the clinical indication.

Specialized techniques, including Doppler ultrasound, elastography, and contrast-enhanced ultrasound, may also be appropriate in selected circumstances. These techniques serve different purposes and should not automatically be considered necessary for every examination.

The important question is whether the examination performed was appropriate for the intended purpose and whether its scope was communicated accurately.

For example, an examination focused on the gallbladder and biliary system should not automatically be interpreted as a comprehensive assessment of every abdominal organ. Similarly, focused imaging of a previously identified lesion may be sufficient for a particular follow-up question without replacing a broader examination when one is clinically indicated.

This distinction also matters when comparing comprehensive diagnostic ultrasound with point-of-care ultrasound (POCUS).

Focused bedside ultrasound may be performed to answer a specific clinical question within a defined scope. Its protocols, documentation, interpretation, and professional responsibilities should reflect the clinical purpose and setting. They should not automatically be assumed identical to those of a comprehensive diagnostic imaging service.

The AIUM's 2026 abdominal and retroperitoneal practice parameter specifically distinguishes the diagnostic ultrasound examinations addressed by that document from POCUS.

Within a diagnostic ultrasound service, the clinical indication, examination protocol, actual assessment, and final report should be reasonably aligned. When the examination must be modified because of clinical circumstances, relevant variations should be recognized and documented appropriately.

Completeness should therefore not be judged simply by counting the number of organs mentioned in a report or the number of images stored.

An appropriately performed focused examination should not be classified as incomplete because it lacks images expected from a comprehensive study. Conversely, a report containing a standard list of normal findings does not establish that a comprehensive examination was performed.

The reliability of the conclusion depends on whether the available examination evidence supports what the report claims was assessed.

Recognizing Technical Limitations Without Overstating Diagnostic Confidence

Some limitations of ultrasound cannot be eliminated through better technique alone.

Bowel gas may obscure portions of the pancreas or other abdominal structures. Body habitus may limit penetration or resolution. Pain, restricted mobility, or an inability to maintain a suitable position may make certain views difficult to obtain. Anatomical variation and the characteristics of an underlying condition may introduce further uncertainty.

These circumstances do not automatically indicate poor professional performance. The concern arises when a limitation that materially affects interpretation is not recognized or communicated.

A report stating that an organ appears normal may create greater confidence than the examination supports if that organ was not adequately visualized.

Where appropriate, the report should distinguish between structures that were assessed satisfactorily and those for which meaningful evaluation was limited.

The clinical implications depend on the question being investigated. Partial visualization may provide enough information to answer one question but remain insufficient for another. A limitation affecting an anatomical region unrelated to the immediate concern may have different implications from one involving the suspected site of disease.

The responsible interpreting professional must determine whether the information obtained adequately supports the conclusion.

When an examination cannot answer the intended clinical question, further assessment may be appropriate. However, this does not mean that every technically limited ultrasound should automatically be repeated.

Depending on the circumstances, subsequent care may involve another ultrasound examination under modified conditions, an alternative imaging modality, specialist review, or a decision that no additional investigation is currently justified.

These decisions should consider clinical risk, the limitations of the original examination, and the potential benefits and harms of further testing.

The important safeguard is that clinically meaningful uncertainty remains visible rather than being concealed by a report that sounds more definitive than the evidence allows.

Image Documentation: Preserving Meaningful Evidence of the Examination

Ultrasound is a dynamic examination. During scanning, the operator continuously adjusts the transducer's position, orientation, and pressure while observing anatomical structures from different perspectives.

A stored still image captures only a selected moment from that process. Even a series of images may not reproduce everything observed during real-time assessment.

This makes image documentation particularly important.

The AIUM Practice Parameter for Documentation of an Ultrasound Examination, 2025 Revision, first published on January 10, 2026, emphasizes the need for a permanent record of the examination and its interpretation.

The guidance recommends recording relevant normal and abnormal areas in a retrievable format that supports subsequent evaluation of diagnostic quality. It also recognizes that still-frame images may be sufficient for some examinations, while dynamic image sequences or cine loops may be required or preferable for others.

The appropriate approach depends on the examination and its relevant professional parameter.

A static image may adequately demonstrate a measurement or a particular anatomical finding. It may be less suitable for documenting certain dynamic abnormalities or structures that need to be evaluated across multiple scanning planes.

Conversely, storing a large number of cine loops does not automatically demonstrate that the examination was complete or clinically appropriate.

The aim is to preserve sufficient relevant evidence, not simply to maximize the number of stored images.

The AIUM documentation parameter identifies important information that should accompany archived ultrasound images, including patient and facility identification, examination date and time, anatomical location and laterality where appropriate, image orientation, and displayed acoustic output indices.

For digitally stored images, this information may be contained in metadata rather than displayed directly on every image, provided it remains accessible and readable during review.

These details support clinical interpretation, traceability, and reliable linkage between the examination and the correct patient.

An archive containing technically excellent images may still be unreliable if those images cannot be confidently associated with the correct patient, examination, or anatomical location.

Image documentation should also be evaluated in relation to the intended examination. A limited but clinically appropriate study may require different supporting images from a comprehensive examination. Documentation requirements should reflect the relevant clinical purpose and professional guidance rather than an arbitrary target for the number of images saved.

Image Acquisition and Image Availability Are Different Responsibilities

An ultrasound examination may be performed appropriately and its images saved on the device, yet those images may not reach the intended clinical archive.

This creates a separate diagnostic risk.

Modern ultrasound services may depend on connected systems for patient registration, examination scheduling, modality worklists, image acquisition, picture archiving and communication systems (PACS), and electronic medical records.

When these systems do not exchange information reliably, image availability, patient identification, and examination traceability may be affected.

A case report published by Osei-Donkor and colleagues in Frontiers in Health Services on September 3, 2026, illustrates this problem.

The authors analyzed recurrent ultrasound workflow failures involving a hospital in Sweden. Missing examination bookings, bypassing of the modality worklist, and manual entry of patient information contributed to problems with imaging metadata, downstream storage, and image retrieval.

Examinations could be technically completed, but the associated information was not always transferred or linked reliably through the intended digital workflow.

The case documentation identified no confirmed patient harm. However, the observed failures created potential risks involving delayed access to diagnostic information, incorrect patient attribution, and confidentiality.

This was a case-based analysis, not a study estimating the frequency of similar incidents across ultrasound services.

Its relevance lies in demonstrating that successful image acquisition does not guarantee that the resulting information remains accessible and correctly attributed.

For ultrasound services, image acquisition, successful transfer, archival integrity, and subsequent retrieval should therefore be considered separately.

Depending on the organization's information systems, useful safeguards may include reliable patient and examination identification, appropriate modality worklist use, confirmation of successful image transfer, and procedures for investigating missing or mismatched studies.

Exceptional circumstances deserve particular attention. Urgent examinations, interrupted network connections, and situations requiring manual data entry should have appropriately controlled procedures so that studies can be reconciled with the correct patient record.

An imaging archive should not be considered reliable solely because the ultrasound device indicates that images were saved.

The clinically relevant question is whether the images remain available to the professionals who need them, with sufficient confidence in their identity, integrity, and traceability.

Ultrasound Reporting Should Reflect What Was Actually Examined

The final ultrasound report is the principal clinical document communicating the interpreting professional's findings and conclusions.

Its reliability depends on whether it accurately represents the examination and provides information relevant to the clinical question.

The AIUM's revised documentation parameter emphasizes a signed final report containing the examination performed, relevant clinical information, findings, interpretation, and appropriate recommendations.

The report should also account for the failure to include standard views or other necessary components when these omissions are relevant to the examination.

This is particularly important when technical limitations affect diagnostic confidence.

A report should not imply that an anatomical structure was comprehensively evaluated if it was not adequately visualized. Nor should a technically limited examination be described in terms suggesting greater certainty than the evidence supports.

Relevant abnormalities should be communicated using terminology that allows the referring clinician to understand their location, characteristics, and potential significance.

Standardized reporting can improve consistency, but templates require care.

A predefined report may contain statements describing normal anatomical findings. If those statements remain unchanged when the examination was limited or an expected structure could not be assessed, the report may misrepresent what was actually observed.

A completed template is therefore not necessarily evidence of a diagnostically adequate examination.

Healthcare organizations should review the accuracy of templated reporting, including whether standard normal statements are appropriately modified when clinical circumstances require.

The report should also distinguish observed findings from professional interpretation and recommendations for subsequent assessment.

When additional imaging or clinical evaluation is suggested, the recommendation should be proportionate to the findings and limitations.

The objective is not to generate more investigations. It is to communicate information that supports appropriate clinical decisions.

When Preliminary and Final Interpretations Differ

In some clinical situations, ultrasound findings may influence patient care before the final report is completed. A preliminary interpretation may be communicated during urgent assessment or while formal reporting is still underway.

The preliminary and final interpretations should not automatically be assumed identical.

Subsequent review may identify an additional finding, clarify an uncertain observation, or change the significance attributed to the initial examination.

The AIUM documentation parameter recognizes that preliminary reports may influence clinical decisions and should be appropriately documented and preserved when issued. It also emphasizes communication of clinically significant differences between preliminary and final interpretations.

This matters because treatment or disposition decisions may already have been made using preliminary information.

Not every change in wording or minor editorial correction requires urgent communication. The response should reflect the potential clinical significance of the change.

Where a revised interpretation could materially affect patient management, the responsible healthcare professional should receive the updated information through an appropriate communication process.

These arrangements should be consistent with the organization's procedures for urgent findings, amended reports, and communication between clinical services.

Professional Competence in a Real-Time Imaging Environment

Ultrasound places particular demands on the person performing the examination.

The operator needs to understand relevant anatomy, select appropriate imaging settings, recognize artifacts and anatomical variations, and determine whether additional views are necessary.

The ability to obtain an ultrasound image is not necessarily equivalent to the competence required to perform a complete diagnostic examination.

Likewise, the professional interpreting archived images must understand their limitations, including the possibility that relevant observations made during live scanning were not preserved.

The relationship between image acquisition, interpretation, and professional responsibility therefore needs to be clearly established.

In some healthcare settings, an appropriately qualified physician both performs and interprets the examination. In others, image acquisition may be undertaken by appropriately trained personnel within an established supervisory and interpretation model.

The appropriate arrangement depends on professional qualifications, applicable regulations, and the healthcare organization's authorized activities.

The AIUM's Standards and Guidelines for the Accreditation of Ultrasound Practices, updated on April 20, 2026, address training, credentials, professional performance, supervision, and quality assurance within its accreditation program.

These standards emphasize ongoing evaluation of personnel rather than reliance on initial qualifications alone.

For healthcare organizations, the relevant quality improvement principle is that competence should be maintained and assessed in relation to the examinations actually performed.

Experience in routine abdominal ultrasound does not automatically establish competence in specialized vascular, obstetric, musculoskeletal, or other advanced ultrasound applications.

Different examinations may require additional technical knowledge, practical skills, and interpretive capabilities.

Training records, qualifications, and examination volume can support competence assessment, but none independently establishes that every examination is performed reliably.

Reviewing representative examinations, providing feedback on technical limitations, and evaluating reporting quality may offer more meaningful evidence of continuing performance.

Such review should also consider case complexity, equipment capabilities, and the clinical environment in which examinations are performed.

Professional competence is supported not only by individual training but also by appropriate supervision, equipment, and organizational systems.

Reviewing Examination Quality Without Assuming Every Difference Is an Error

Retrospective quality review can help healthcare organizations understand whether ultrasound examinations consistently meet their intended clinical purpose.

However, reviewing the final report alone may not provide enough information.

Where available, reviewers may need to examine archived images, assess whether relevant anatomical components were documented, and determine whether the report accurately represents the findings.

The AIUM accreditation standards updated in April 2026 require participating practices to demonstrate ongoing evaluation of ultrasound personnel through regular retrospective performance reviews. These requirements address technical performance, documentation, reporting, and other aspects of service quality.

They apply within the relevant AIUM accreditation framework and should not automatically be treated as legal obligations for every healthcare facility in Viet Nam.

For local quality improvement, structured reviews can help distinguish different concerns.

An examination may have inadequate image quality because of unsuitable equipment settings or difficult acoustic conditions. Another may contain technically satisfactory images but insufficient documentation of an expected anatomical component.

A report may describe a finding inaccurately even when the archived images are adequate. Alternatively, an examination may have been appropriately performed, but relevant images were not successfully transferred to the archive.

These problems arise at different stages and should not receive identical corrective responses.

Retrospective review also has limitations.

An archived image demonstrates what was saved, not necessarily everything observed during live scanning.

The absence of an expected image may establish a documentation gap without proving that the anatomical structure was never examined. Conversely, the presence of a normal image does not establish that the entire relevant region was adequately assessed.

Reviewers should distinguish what the available evidence demonstrates from what can only be inferred.

This distinction is particularly important when a later clinical diagnosis becomes known.

A subsequent diagnosis does not automatically prove that the earlier ultrasound was incorrectly performed or interpreted. The patient's condition may have changed, new findings may have developed, or later investigations may have provided information unavailable during the original examination.

A fair quality review should consider the clinical information, technical limitations, and professional expectations applicable at the time.

The purpose is to identify credible opportunities for improvement without turning every subsequent diagnosis into an allegation of error.

Peer Review and Learning From Diagnostic Discrepancies

Peer review allows appropriately qualified professionals to evaluate selected examinations and discuss differences in image acquisition, interpretation, or reporting.

Some discrepancies concern technical adequacy, while others reflect differences in how findings are interpreted.

Two professionals may reach different conclusions from the same ultrasound images without either conclusion necessarily representing a confirmed error.

Ultrasound interpretation may be influenced by image quality, anatomical variation, the dynamic nature of scanning, and the clinical information available.

When differences are clinically significant, further review may be appropriate. This might involve professional discussion, consultation with another qualified specialist, or consideration of additional diagnostic evidence.

Relevant evidence may include subsequent imaging, clinical assessment, laboratory findings, surgical observations, or pathology results, depending on the condition involved.

However, later investigations should not automatically be treated as perfect reference standards. They may examine different aspects of disease, be performed under different conditions, or reflect changes occurring after the original study.

Clinical correlation therefore requires judgment.

A useful review process should identify whether the concern involved image acquisition, recognition of a finding, interpretation, documentation, or information transfer. It should also consider whether similar concerns recur.

Repeated difficulty obtaining particular views may suggest a need for targeted training or protocol adjustment. Frequent labeling problems may indicate a workflow or user-interface weakness. Recurring differences in reporting terminology may benefit from professional discussion and appropriate standardization.

Peer review should encourage learning and accountability without promoting defensive reporting or unnecessary investigations.

Where a finding has potential implications for a patient's current care, the review process should also support appropriate clinical communication and action.

Equipment Quality and the Reliability of Diagnostic Images

Diagnostic ultrasound depends on equipment that performs appropriately for its intended clinical use.

Ultrasound systems vary in transducer capabilities, imaging software, resolution, Doppler functions, and other technical features. Equipment suitable for one clinical application may not be adequate for another.

Selection and maintenance should therefore reflect the examinations performed and the capabilities needed to obtain clinically useful information.

Routine quality assurance is an important part of this responsibility.

Transducer damage, malfunctioning elements, image artifacts, and changes in system performance can affect image quality and potentially influence interpretation. Some problems may develop gradually and may not be immediately obvious during routine clinical use.

The AIUM accreditation standards updated in April 2026 require participating practices to perform documented ultrasound equipment quality assurance testing at least annually, or more frequently when concerns arise.

The standards also refer to Routine Quality Assurance of Clinical Ultrasound Equipment: Version 2.0.

This annual frequency is an AIUM accreditation requirement. It should not automatically be presented as a universal statutory inspection interval for ultrasound equipment in Viet Nam.

Healthcare organizations should establish monitoring, maintenance, and technical testing arrangements consistent with applicable Vietnamese requirements, manufacturer instructions, clinical risks, and relevant accreditation commitments.

Equipment quality assurance may include inspection of transducers, evaluation of image uniformity, assessment of suspected artifacts, and other tests appropriate to the equipment and its intended use.

When a defect is identified, the organization should consider whether it may have affected examinations performed before the problem was discovered.

For example, a damaged transducer may require repair or replacement. Depending on the nature of the defect and potential consequences, selected earlier examinations may also warrant review.

The need for retrospective assessment should be determined by the available evidence and clinical risk rather than assuming that every examination performed with the equipment was unreliable.

A completed maintenance checklist does not establish that every subsequent image will be diagnostically adequate.

Equipment quality assurance and professional examination review are complementary safeguards.

Ultrasound Safety Beyond Image Quality

Diagnostic ultrasound does not involve ionizing radiation, but this does not mean that safety considerations can be ignored.

The AIUM's As Low As Reasonably Achievable (ALARA) Principle, reaffirmed on August 17, 2026, emphasizes appropriate use of acoustic exposure while obtaining the information needed for diagnosis.

The guidance addresses examination presets, acoustic power, exposure duration, and monitoring of mechanical and thermal indices.

Particular care may be necessary for examinations involving sensitive tissues or circumstances in which acoustic exposure requires closer attention.

ALARA should not be interpreted as reducing output to a level that makes the examination diagnostically inadequate. The principle is to obtain necessary clinical information while minimizing unnecessary exposure.

Infection prevention is another important component of safe ultrasound practice.

Transducers may contact intact skin, mucous membranes, or tissues involved in invasive procedures. The appropriate level of cleaning and disinfection depends on how the equipment is used.

The AIUM's July 2025 guidance on cleaning and preparing ultrasound transducers distinguishes external examinations from internal and interventional applications.

External transducers used on intact skin generally require cleaning and appropriate low-level disinfection between patients. Internal transducers require more stringent reprocessing, including high-level disinfection and appropriate single-use protective covers.

Requirements for invasive procedures depend on the nature of the intervention and the level of sterility needed.

Healthcare organizations should also consider manufacturer instructions, applicable infection prevention requirements, and appropriate ultrasound gel handling.

These safeguards are necessary for patient safety, although they address different risks from examination completeness and diagnostic interpretation.

A reliable ultrasound service must consider both the quality of the diagnostic information produced and the safety of the process used to obtain it.

Measuring Ultrasound Quality Without Misleading Indicators

Quality measurement can help healthcare organizations identify problems that may not be apparent from individual examinations.

However, diagnostic ultrasound quality cannot be represented adequately by a single percentage.

A meaningful measurement program should begin by defining the types of examinations being evaluated.

Abdominal, thyroid, obstetric, vascular, and musculoskeletal ultrasound examinations have different technical requirements and clinical purposes. The indicators used should reflect the relevant examination categories and applicable protocols.

One useful measure may be the proportion of eligible examinations for which appropriate anatomical documentation is available. Another may assess whether reports clearly identify technical limitations that materially affect interpretation.

Organizations may also monitor patient and examination identification accuracy, image archival and retrieval problems, the frequency of structured quality review, and selected technical concerns identified through peer assessment.

Repeat examinations may provide useful information, but their interpretation requires care.

A high repeat-examination rate may reflect technical inadequacy, but it may also result from appropriate further assessment of complex cases or changes in the patient's condition.

A low repeat rate is not necessarily evidence of superior quality if inadequate examinations are not recognized or appropriate repeat studies are not requested.

Similarly, frequent documentation of technical limitations does not automatically indicate poor performance. It may reflect a challenging patient population or a reporting culture that accurately recognizes uncertainty.

The more meaningful concern is whether limitations were appropriately identified, communicated, and managed.

Denominators should also reflect the examinations being assessed.

A measure of examination completeness should apply to cases for which the specified protocol components were relevant and reasonably assessable.

A quality review based only on archived studies may fail to identify examinations whose images were not stored successfully.

Peer review focused exclusively on difficult or disputed cases may identify important risks but cannot provide an unbiased estimate of overall service performance.

Combining targeted review of higher-risk cases with a systematic sample of routine examinations can provide a more balanced understanding.

Measurement should also distinguish process quality from clinical outcomes.

Better documentation does not, by itself, prove that fewer abnormalities were missed. A low discrepancy rate does not necessarily establish that interpretation is more reliable if the review process lacks sufficient coverage or sensitivity.

The objective is to identify credible improvement opportunities, evaluate whether interventions address the intended problems, and understand whether the ultrasound service is becoming more dependable.

Building a Practical Ultrasound Quality Review Program

Healthcare organizations can begin improving ultrasound reliability without immediately launching an extensive accreditation project.

A practical approach is to select one defined ultrasound service and examine how its current examinations are performed, documented, reported, and stored.

For example, a hospital may begin with routine abdominal ultrasound because the examination is frequently performed and a substantial number of archived studies may already be available.

The organization should first determine which examination types are included and identify the relevant professional parameters or locally approved protocols.

A sample of examinations can then be reviewed to assess whether the clinical indication, intended scope, anatomical documentation, and final report are reasonably aligned.

The review should consider technical adequacy, relevant image documentation, reporting accuracy, and recognition of material limitations. It should also establish whether images can be retrieved and confidently linked to the correct patient and examination.

Different services may reveal different weaknesses.

One may find that examinations are generally technically adequate but image labeling is inconsistent. Another may identify recurring difficulties visualizing particular anatomical structures. A third may discover that examination reports are complete but images are not reliably accessible in the archive.

These problems require different responses.

Documentation weaknesses may justify changes to image capture or labeling procedures. Acquisition problems may require targeted technical training, equipment adjustments, or supervision. Reporting concerns may benefit from clearer terminology or more appropriate templates.

Image storage problems may require investigation of patient identifiers, modality worklists, system interfaces, and archive workflows.

The review should involve professionals who perform and interpret ultrasound, supported where appropriate by quality personnel, information technology teams, and biomedical engineering staff.

Their involvement matters because some causes of poor performance may be difficult to recognize from reports alone.

After implementing changes, the organization should review another sample of examinations to determine whether the identified weakness has improved.

The purpose is not to produce more policies, collect more images, or increase administrative work. It is to improve the reliability of the examination and the information available for clinical decisions.

Diagnostic Ultrasound Within Viet Nam's Legal Framework

Diagnostic ultrasound services in Viet Nam operate within the broader legal framework governing medical examination and treatment, professional practice, healthcare facility activities, medical records, and medical devices.

The Law on Medical Examination and Treatment No. 15/2023/QH15, as amended and reflected in Consolidated Document No. 26/VBHN-VPQH dated February 26, 2026, provides an important legal foundation.

Healthcare facilities and practitioners must perform medical activities within applicable licensing arrangements, authorized professional scope, and relevant technical requirements.

These provisions are relevant when organizations establish who may perform, interpret, and report ultrasound examinations.

However, international practice parameters should not be interpreted as independently authorizing particular professional groups to perform or interpret examinations in Viet Nam.

The appropriate responsibilities depend on Vietnamese law, professional qualifications, authorized activities, and relevant organizational arrangements.

Another important reference is Circular No. 23/2024/TT-BYT, issued by the Ministry of Health on October 18, 2024, concerning the list of medical examination and treatment techniques.

Circular No. 25/2026/TT-BYT, issued on June 30, 2026, amended certain provisions, including those relating to Circular No. 23/2024/TT-BYT, with the relevant amendments taking effect on August 15, 2026.

The consolidated technical list is reflected in Consolidated Document No. 13/2026/VBHN-TT-BYT, dated July 30, 2026.

These instruments are relevant when healthcare organizations determine which technical procedures fall within their authorized services and how those activities should be organized.

However, inclusion of a procedure in a technical list does not establish that every individual examination is technically adequate or diagnostically reliable.

Likewise, AIUM accreditation standards and practice parameters do not automatically become mandatory under Vietnamese law.

AIUM states that its ultrasound practice parameters and accreditation framework are not intended to establish a legal standard of care.

Healthcare organizations may use these international resources to inform quality improvement while ensuring that local procedures remain consistent with Vietnamese requirements and the actual scope of services provided.

Medical Device Management, Records, and Patient Data

Ultrasound equipment is subject to relevant requirements under Viet Nam's medical device regulatory framework.

Decree No. 98/2021/ND-CP on medical device management, as amended and reflected in Consolidated Document No. 08/VBHN-BYT dated March 6, 2026, provides an important reference.

Healthcare organizations should consider applicable requirements concerning the equipment they procure and use, its intended purpose, supporting documentation, and relevant management obligations.

Compliance with medical device requirements, however, does not automatically establish the quality of every diagnostic examination performed using that equipment.

The organization needs both appropriately managed devices and qualified personnel working within reliable clinical procedures.

Medical records are another important consideration.

Article 69 of the Law on Medical Examination and Treatment establishes requirements concerning medical records, including their creation, updating, preservation, confidentiality, and access.

These requirements are relevant to ultrasound findings, diagnostic reports, and associated examination information.

Circular No. 13/2025/TT-BYT, issued on June 6, 2025, provides guidance on implementing electronic medical records in healthcare facilities and took effect on July 21, 2025.

Electronic medical records and associated imaging systems can support the availability and traceability of diagnostic information.

However, implementing an electronic medical record does not automatically establish that every ultrasound examination has adequate image documentation or that all archived studies can be retrieved reliably.

These capabilities require appropriate technical arrangements, workflow controls, and ongoing evaluation.

Healthcare organizations should also distinguish international professional recommendations on image documentation from specific record retention requirements established by Vietnamese law and applicable institutional policies.

Retention, access, and disposal arrangements should reflect applicable requirements and legitimate clinical needs.

Protecting personal information is equally important.

Viet Nam's Law on Personal Data Protection No. 91/2025/QH15 and Decree No. 356/2025/ND-CP took effect on January 1, 2026.

Ultrasound records may contain identifiable patient information, clinical indications, diagnostic findings, images, and examination metadata.

Healthcare organizations should assess applicable requirements concerning lawful processing, confidentiality, authorized access, information security, and disclosure or transfer of personal data.

Consent requirements, applicable exceptions, and other legal conditions depend on the nature and purpose of the processing activity.

When ultrasound images are transferred between facilities, stored through external systems, or accessed for quality review, appropriate safeguards should protect patient information.

Internal quality assessment, external professional consultation, research, and educational use may involve different purposes and should not automatically be treated as legally interchangeable.

A dependable ultrasound service must preserve clinically important diagnostic information while protecting the confidentiality and rights of the people to whom that information relates.

From Image Acquisition to Dependable Diagnostic Information

Diagnostic ultrasound provides valuable information across many areas of healthcare, but its reliability depends on what happens during the examination as well as what appears in the final report.

The person performing the examination must obtain appropriate views, recognize relevant findings, and understand the limitations of the technique. The interpreting professional must determine whether the available information supports the conclusion and communicate material uncertainty when necessary.

The healthcare organization must ensure that relevant images are preserved, equipment is appropriately maintained, and quality concerns can be reviewed constructively.

The AIUM documentation guidance published in January 2026, its accreditation standards updated in April, and the revised abdominal ultrasound parameter published in September provide current international references for these responsibilities.

For healthcare organizations in Viet Nam, the priority should be to develop practical quality review processes aligned with the examinations performed, the competence of authorized professionals, and the clinical risks of the service.

This does not mean every examination must follow an identical scanning protocol, every report must contain the same number of images, or every technical limitation must lead to another investigation.

It means that the intended scope should be clear, the findings and limitations should be represented accurately, and sufficient relevant evidence should remain available for interpretation and subsequent review.

Diagnostic ultrasound reliability is not demonstrated merely by acquiring images or issuing a report. It depends on whether the examination adequately addresses the clinical question, whether the findings and limitations are documented faithfully, and whether the healthcare organization can evaluate and improve the quality of the information on which clinical decisions depend.

Selected References and Regulatory Sources

American Institute of Ultrasound in Medicine. Standards and Guidelines for the Accreditation of Ultrasound Practices. Updated April 20, 2026.

American Institute of Ultrasound in Medicine. AIUM Practice Parameter for Documentation of an Ultrasound Examination, 2025 Revision. Journal of Ultrasound in Medicine. 2026;45(6)–E29. First published January 10, 2026. DOI: 10.1002/jum.70168.

American Institute of Ultrasound in Medicine. AIUM Practice Parameter for the Performance of an Ultrasound Examination of the Abdomen and/or Retroperitoneum, 2026 Revision. Journal of Ultrasound in Medicine. Published online September 11, 2026. DOI: 10.1002/jum.70430.

American Institute of Ultrasound in Medicine. Supervision and Interpretation of Ultrasound Examinations. Reapproved April 20, 2026.

American Institute of Ultrasound in Medicine. As Low As Reasonably Achievable (ALARA) Principle. Reapproved August 17, 2026.

American Institute of Ultrasound in Medicine. Guidelines for Cleaning and Preparing External- and Internal-Use Ultrasound Transducers and Equipment Between Patients as Well as Safe Handling and Use of Ultrasound Coupling Gel, 2025 Revision. July 2, 2025.

American Institute of Ultrasound in Medicine. Routine Quality Assurance of Clinical Ultrasound Equipment. Version 2.0.

Society of Radiographers and British Medical Ultrasound Society. Guidelines for Professional Ultrasound Practice. Ninth edition, Revision 8. December 2025.

Osei-Donkor B, Hassoun A, Jabin MSR. Patient Misidentification and Data Integrity Risks in Ultrasound Imaging: A Case Report Using Multi-Framework Incident Analysis. Frontiers in Health Services. 2026;6:1872556. Published September 3, 2026. DOI: 10.3389/frhs.2026.1872556.

National Assembly of Viet Nam. Law on Medical Examination and Treatment No. 15/2023/QH15, as amended. See also Consolidated Document No. 26/VBHN-VPQH, dated February 26, 2026.

Ministry of Health of Viet Nam. Circular No. 23/2024/TT-BYT, Promulgating the List of Medical Examination and Treatment Techniques. October 18, 2024, as amended.

Ministry of Health of Viet Nam. Circular No. 25/2026/TT-BYT, Amending and Supplementing Certain Provisions of Ministry of Health Circulars, Including Circular No. 23/2024/TT-BYT. June 30, 2026. Effective August 15, 2026.

Ministry of Health of Viet Nam. Consolidated Document No. 13/2026/VBHN-TT-BYT, Concerning the List of Medical Examination and Treatment Techniques. July 30, 2026.

Government of Viet Nam. Decree No. 98/2021/ND-CP on Medical Device Management, as amended and reflected in Consolidated Document No. 08/VBHN-BYT dated March 6, 2026.

Ministry of Health of Viet Nam. Circular No. 13/2025/TT-BYT, Providing Guidance on the Implementation of Electronic Medical Records in Healthcare Facilities. June 6, 2025. Effective July 21, 2025.

National Assembly of Viet Nam. Law on Personal Data Protection No. 91/2025/QH15. Effective January 1, 2026.

Government of Viet Nam. Decree No. 356/2025/ND-CP, Detailing Certain Provisions and Measures for Implementing the Law on Personal Data Protection. Effective January 1, 2026.

Professional Scope and Disclaimer

This article is published by Diagnostic Excellence Vietnam (DxEx.vn), a professional initiative dedicated to advancing diagnostic quality, safety, and reliability in Viet Nam through evidence-informed knowledge sharing, risk management, measurement, and quality improvement.

The content is intended for general professional education and discussion. It does not constitute patient-specific medical advice, a clinical or imaging diagnosis, an individualized recommendation concerning ultrasound examination techniques, image interpretation, further investigations, treatment, or follow-up, a determination of professional or technical error, an accreditation assessment, or legal advice.

References to international professional guidelines, practice parameters, technical standards, and accreditation requirements do not imply that these recommendations are automatically mandatory under Vietnamese law. Their applicability depends on relevant legal requirements, professional scope, examination type, healthcare setting, and organizational circumstances.

Decisions concerning the selection and performance of ultrasound examinations, professional competence, image acquisition, interpretation, reporting, equipment quality assurance, clinical communication, and patient management must be made by appropriately qualified and authorized healthcare professionals in accordance with applicable Vietnamese law, relevant professional standards, authorized service capabilities, and the patient's clinical circumstances.

Diagnostic Excellence Vietnam does not provide ultrasound examinations, clinical diagnosis, or treatment services through this educational publication.